FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3093822 · Received May 6, 2013

Report

Report Number
3004209178-2013-93137
Event Type
Malfunction
Date Received
May 6, 2013
Date of Event
April 9, 2013
Report Date
April 9, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RESERVOIR LEAKED INSULIN PAST BOTH O-RINGS DURING NORMAL USE INTO THE INSULIN PUMP RESERVOIR COMPARTMENT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197003 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 3 YR