ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-05903
- Event Type
- Malfunction
- Date Received
- May 6, 2013
- Date of Event
- March 22, 2013
- Report Date
- April 12, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1 DATE OF SUBMISSION 08/06/2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT. EVALUATION REVEALED THAT THE UP ARROW KEYPAD BUTTON WAS UNRESPONSIVE. THERE WAS NO EVIDENCE OF CONTAMINATION FOUND UNDER THE KEY CONTACTS. THE PUMP COVER WAS REMOVED AND CORROSION WAS FOUND ON THE KEYPAD FLEX. UNRELATED TO THE COMPLAINT, MOISTURE AND CORROSION WAS NOTED ON COMPONENTS INSIDE THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE DISTRIBUTOR REPORTED THE UP, DOWN, AND OK BUTTONS WERE UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 197438 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |