FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3093563
·
Received May 6, 2013
Report
- Report Number
- 2531779-2013-05895
- Event Type
- Malfunction
- Date Received
- May 6, 2013
- Report Date
- April 8, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED CONTAMINATION TO THE UP ARROW AND OK KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION TO THE UP ARROW AND OK KEY CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.