FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3093545 · Received May 6, 2013

Report

Report Number
2531779-2013-05893
Event Type
Malfunction
Date Received
May 6, 2013
Report Date
April 8, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE TEXT ON THE DISPLAY SCREEN IS DIM FADED AND DISCOLORED UPON START UP AND DIFFICULT TO READ. THE DISPLAY LENS IS HEAVILY SCRATCHED AND DIFFICULT TO READ. EVALUATION WAS COMPLETED WITH A TEST DISPLAY SCREEN, THE TEST SCREEN IS FULLY FUNCTIONAL AND FULLY ILLUMINATED WITH NO VISIBLE SIGNS OF FADING. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE KEYPAD SYMBOLS WERE WORN, WHICH HAS NO EFFECT ON INSULIN DELIVERY. (B)(6). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RE/CORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM AND SCRATCHED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
197246 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 60 YR