FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 3093363 · Received April 26, 2013

Report

Report Number
1828100-2013-00461
Event Type
Malfunction
Date Received
April 26, 2013
Date of Event
March 28, 2013
Report Date
March 28, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REPRESENTATIVE (FSR). THIS COOLER HEATER UNIT WAS IN STORAGE OFF-SITE SINCE 2009. THE SCALE AND DEBRIS BUILD-UP WAS THIS UNIT WAS EXTREME. THE FSR CLEANED THE UNIT BY FLUSHING OUT WITH WATER, WHICH WAS VERY WEAK. THE FSR DE-SCALED THE COOLER HEATER PER THE UNITS GUIDELINES. THIS WAS VERY TIME CONSUMING FOR MULTIPLE DE-SCALINGS ON THE UNIT. WITH THE DE-SCALINGS AND PREVENTIVE MAINTENANCE COMPLETED, THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, WATER WAS FLOWING VERY WEAK THROUGH THE COOLER HEATER UNIT. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
181684 TERUMO TCM II COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4415

Patients

Seq Age Sex Outcome Treatment
1