ACCU-CHEK SPIRIT COMBO
Report
- Report Number
- 2183996-2013-00788
- Event Type
- Malfunction
- Date Received
- May 6, 2013
- Date of Event
- April 23, 2013
- Report Date
- July 11, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
CONCLUSION THE COMPLAINT CAN BE VERIFIED. RESULT THE MENU BUTTON DOES NOT RESPOND. THERE ARE PARTICLES INSIDE THE HOUSING OF THE MENU BUTTON. THE PARTICLES ISOLATE THE AREA OF CONTACT INSIDE THE BUTTON HOUSING. DUE TO THIS PROBLEM THE MENU BUTTON DOES NOT RESPOND AND IS WITHOUT FUNCTION.
THIS INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.
PATIENT REPORTED A NON-FUNCTIONAL MENU BUTTON ON THE INFUSION DEVICE. NO FURTHER INFORMATION IS AVAILABLE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. REQUESTED RETURN OF THE ALLEGED INFUSION DEVICE FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 196561 | ACCU-CHEK SPIRIT COMBO | INSULIN INFUSION PUMP | LZG | ROCHE DIABETES CARE AG | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 054 YR |