COULTER® LH 500 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2013-00748
- Event Type
- Malfunction
- Date Received
- May 6, 2013
- Date of Event
- April 10, 2013
- Report Date
- April 10, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K032000
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE FSE INSPECTED THE INSTRUMENT AND CONFIRMED A CONTAINED CLENZ LEAK. THE FSE REPLACED A WORN TUBING AT PINCH VALVE (PV37), RESOLVING THE LEAK. THE FSE COMPLETED A FLOW CELL ALIGNMENT TO RESOLVE THE LATRON CONTROL RECOVERY. THE FAILURE MODE WAS RELATED TO WORN TUBING THROUGH PV37 AND TO ALIGNMENT OF THE FLOW CELL WHICH CAUSED THE LATRON CONTROL TO BE OUT FOR THE VOLUME CHANNEL. (B)(4).
THE CUSTOMER REPORTED TO BECKMAN COULTER THAT LESS THAN 2 ML OF CLENZ WAS LEAKING FROM THE RIGHT HAND SIDE DILUTER CENTER PANEL OF THE COULTER LH 500 HEMATOLOGY ANALYZER. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE FLUID HAD LEAKED ONTO THE TRIPLE TRANSDUCER MODULE (TTM) PLASTIC COVER. THE CUSTOMER ALSO REPORTED THE LATRON VOLUME WAS LOW AT 24.1. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED BY THE CUSTOMER. THERE IS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE AT THE FACILITY. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE), CONSISTING OF GLOVES, FACE PROTECTION AND A LABORATORY COAT WHEN THE LEAK WAS OBSERVED. THERE WAS NO BIOHAZARD EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. THERE WAS NO IMPACT TO PATIENT RESULTS. NO DEATH OR INJURY IS ATTRIBUTED TO THIS EVENT AND THERE WAS NO CHANGE TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 196084 | COULTER® LH 500 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | LH 500 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |