FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600I SYNCHRON® ACCESS CLINICAL SYSTEM

MDR report key: 3093213 · Received May 5, 2013

Report

Report Number
2050012-2013-00301
Event Type
Malfunction
Date Received
May 5, 2013
Date of Event
April 8, 2013
Report Date
April 8, 2013
Manufacturer
BECKMAN COULTER
Product Code
JJE
PMA / PMN Number
K060256
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER FSE (FIELD SERVICE ENGINEER) VISITED THE CUSTOMER'S LABORATORY MULTIPLE TIMES AND REPLACED MULTIPLE PARTS INVOLVED IN SAMPLE HANDLING. THE FSE NOTED ISSUES WITH SLIGHT OBSTRUCTION IN CC SAMPLE PROBE CAUSING RANDOM SHORT SAMPLES, CC SAMPLE MOTION ERRORS, AND INTERMITTENT LOW SAMPLE VOLUME DRAWN BY CC SYRINGE ON THE INSTRUMENT. THE FSE REPLACED BOTH THE CC SAMPLE PROBE AND AUTOGLOSS WICK ON (B)(4) 2013 BUT THE ISSUE WAS NOT RESOLVED. THE FSE RETURNED THE FOLLOWING DAY ON (B)(4) 2013 TO REPLACE THE CC SAMPLE SYRINGE DRIVE ASSEMBLY AND PERFORMED ALIGNMENTS. FAILURE MODE IS ATTRIBUTED TO THE CC SYRINGE DRIVE THAT THE FSE REPLACED ON (B)(4) 2013. RESULTS: CC SYRINGE DRIVE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED OBTAINING SUPPRESSED (NON-NUMERIC), OUT OF INSTRUMENT RANGE LOW (OIR LO), CC (CARTRIDGE CHEMISTRIES) RESULTS FOR SEVERAL SAMPLES INVOLVING THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM. THE CUSTOMER PROVIDED RESULT PRINTOUTS FOR FOUR (4) PATIENTS. THE CUSTOMER DID NOT REPORT THE INCORRECT RESULTS OUT OF THE LABORATORY AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. THE CUSTOMER DISCOVERED THAT THE CC SAMPLE SYRINGE WAS LOOSE BUT TIGHTENING THE CC SAMPLE SYRINGE DID NOT RESOLVE THE ISSUE. THE CUSTOMER STATED THAT QC (QUALITY CONTROL) RESULTS WERE WITHIN SPECIFICATIONS BEFORE AND AFTER THE EVENT BUT DID NOT PROVIDE QC DATA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195937 UNICEL® DXC 600I SYNCHRON® ACCESS CLINICAL SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1