FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3092951 · Received May 3, 2013

Report

Report Number
2032227-2013-01647
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 7, 2013
Report Date
April 8, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING REWIND DUE TO ENCODER OUT OF PHASE. UNABLE TO PERFORM DISPLACEMENT TEST DUE TO CONSTANT MOTOR ERROR ALARM. UNABLE TO PERFORM BASIC OCCLUSION TEST, OCCLUSION TEST AND EXCESSIVE NO DELIVERY TEST DUE TO CONSTANT MOTOR ERROR ALARM. DRIVE SUPPORT DISK WAS INSPECTED AND NO ANOMALY WAS NOTED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT THAT THE DRIVE SUPPORT CAP WAS PROTRUDING. THE CUSTOMER ALSO REPORTED A MOTOR ERROR ALARM. THE REPORTED BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 299 MG/DL. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195210 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAH

Patients

Seq Age Sex Outcome Treatment
1 42 YR