FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3092948 · Received May 3, 2013

Report

Report Number
1416980-2013-11322
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 26, 2013
Report Date
April 26, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER'S SERVICE CENTER REGARDING A SYSTEM ERROR 2240 (AIR IN TUBING) WHICH OCCURRED ON THE HOMECHOICE (HC) DURING DWELL 4 OF 5 WHILE THE HOME PATIENT (HP) WAS CONNECTED. THERE WAS NOTHING FOUND DURING TROUBLE SHOOTING THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE ALARM. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM AND HELPED THE HP CLEAR IT BY TURNING THE HC OFF AND ON. THE HC THEN ALARMED SYSTEM ERROR 2367, SO THE HP CYCLED THE POWER AGAIN. THE HC WENT BACK TO "PRESS GO TO START." THE HP WOULD FINISH WITH A MANUAL BAG. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195209 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 49 HOMECHOICE