FDA Adverse Event Malfunction Summary report: N

UV FLASH TRANSFER SETS

MDR report key: 3092828 · Received May 3, 2013

Report

Report Number
1416980-2013-11302
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 25, 2013
Report Date
April 25, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: AN EVALUATION OF THE ACTUAL SAMPLE WAS CONDUCTED. VISUAL INSPECTION, LEAK TESTING AND CLEAR PASSAGE TESTING WAS PERFORMED WITH NO ISSUES NOTED. FUNCTIONALLY HAND TIGHTENED A LAB TITANIUM ADAPTER TO THE SET WITHOUT DIFFICULTY NOTED. THE RETURNED SAMPLE WENT THROUGH UNDER-WATER PRESSURE TESTED AND PERFORMED WITHIN SPECIFICATION. THE EVALUATION OF THE RETURNED SAMPLE COULD NOT DUPLICATE THE ALLEGED ISSUE. AS A RESULT, THE CAUSE COULD NOT BE DETERMINED. A REVIEW OF ALL BATCH RECORD DOCUMENTS FOR POTENTIALLY ASSOCIATED LOT NUMBER H12J17075 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE HAS BEEN REPORTED TO BE AVAILABLE FOR EVALUATION AND HAS BEEN REQUESTED. HOWEVER, THE SAMPLE HAS NOT BEEN RECEIVED FOR EVALUATION AS OF YET. A FOLLOW-UP REPORT WILL BE FILED IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED. (SAME PATIENT, EVENT AS IN (B)(4) FOR THE TITANIUM ADAPTOR.)

Description of Event or Problem · 1

A NURSE REPORTED THAT A CONNECTION ISSUE OCCURRED WHEN A UV FLASH TRANSFER SET WAS BEING CHANGED. THE PATIENT CONNECTOR DID NOT CONNECT WELL WITH THE TITANIUM ADAPTOR. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194251 UV FLASH TRANSFER SETS SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H12J17075

Patients

Seq Age Sex Outcome Treatment
1 TITANIUM ADAPTER