ARMADA 14 PTA CATHETER
Report
- Report Number
- 2024168-2013-02781
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- April 10, 2013
- Report Date
- April 11, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LIT
- PMA / PMN Number
- K102705
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE LEAK FROM THE DISTAL TIP OF THE BALLOON COULD NOT BE CONFIRMED; HOWEVER, A LEAK WAS OBSERVED FROM THE GUIDE WIRE PORT OF THE HUB. WHILE A SEARCH OF THE LOT HISTORY RECORD FOR THIS SPECIFIC LOT INDICATED NO RELATED NON-CONFORMANCE RECORDS, BASED ON AN EXPANDED INVESTIGATION, A PRODUCT ISSUE RELATED TO LEAKAGE AT THE HUB WAS IDENTIFIED. THE PROBABLE CAUSE OF THE HUB LEAKAGE WAS RELATED TO THE ADHESIVE MATERIAL USED DURING THE COMPONENT BONDING PROCESS. FURTHER ASSESSMENT OF THIS ISSUE PER SITE OPERATING PROCEDURES WAS PERFORMED. CORRECTIVE AND PREVENTIVE ACTIONS TO ADDRESS THIS ISSUE ARE IN THE PROCESS OF BEING IMPLEMENTED AND THE PERFORMANCE OF THESE DEVICES WILL CONTINUE TO BE MONITORED.
(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION. THE ADDITIONAL ARMADA 14 REFERENCED IS BEING FILED UNDER A SEPARATE MEDWATCH REPORT.
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A LESION LOCATED IN THE ANTERIOR TIBIAL ARTERY, BELOW THE BIFURCATION, THAT WAS NARROW, MILDLY CALCIFIED AND 90% STENOSED. THE ARMADA 1.5 X 120MM X 90CM BALLOON WAS PLACED IN THE TARGET LESION; HOWEVER, THE BALLOON DID NOT INFLATE AT ALL AND CONTRAST MEDIA WAS OBSERVED LEAKING OUT OF THE DISTAL TIP OF THE BALLOON CATHETER. THE DEVICE WAS REMOVED FROM THE ANATOMY WITHOUT ISSUE. THE ARMADA 2.0 X 200MM X 90CM WAS THEN PLACED IN THE TARGET LESION; HOWEVER, THE BALLOON DID NOT INFLATE AT ALL AND AGAIN CONTRAST MEDIA WAS OBSERVED LEAKING OUT OF THE DISTAL TIP OF THE BALLOON CATHETER. THE DEVICE WAS REMOVED FROM THE ANATOMY WITHOUT ISSUE. ANOTHER ARMADA BALLOON CATHETER (UNKNOWN SIZE) WAS USED SUCCESSFULLY FOR TREATMENT WITH A SATISFACTORY RESULT. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR ADVERSE PATIENT EFFECTS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193901 | ARMADA 14 PTA CATHETER | PERIPHERAL DILATATION CATHETER | LIT | AV-TEMECULA-CT | 724426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GUIDE WIRE: HT COMM AND SHEATH: CORDIS |