FDA Adverse Event Malfunction Summary report: N

SOLETRA

MDR report key: 3092781 · Received May 3, 2013

Report

Report Number
3004209178-2013-07302
Event Type
Malfunction
Date Received
May 3, 2013
Report Date
April 22, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
MHY
PMA / PMN Number
P960009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

CONCOMITANT MEDICAL PRODUCTS: PRODUCT ID 7426, SERIAL# (B)(4), IMPLANTED: (B)(6) 2010. PRODUCT TYPE: IMPLANTABLE NEUROSTIMULATOR: PRODUCT ID 3387-40, LOT# J0230913V, IMPLANTED: (B)(6) 2003. PRODUCT TYPE: LEAD: PRODUCT ID 748251, SERIAL# (B)(4), IMPLANTED: (B)(6) 2003. PRODUCT TYPE: EXTENSION: PRODUCT ID 3387-40, LOT# J0230913V, IMPLANTED: (B)(6) 2003. PRODUCT TYPE: LEAD: PRODUCT ID 748251, SERIAL# (B)(4), IMPLANTED: (B)(6) 2003. PRODUCT TYPE: EXTENSION: PRODUCT ID 7438, SERIAL# (B)(4), IMPLANTED: (B)(6) 2003. PRODUCT TYPE: PROGRAMMER, PATIENT. (B)(4).

Description of Event or Problem · 1

ADDITIONAL INFORMATION STATED THE CAUSE OF THE EVENT WAS UNKNOWN. IT WAS ALSO REPORTED THERE WERE IMPEDANCES GREATER THAN 2000 OHMS ON ELECTRODE COMBINATIONS 1-3 AND 0-3 ON THE LEFT SIDE AND THE THERAPY IMPEDANCE ON THE RIGHT SIDE WAS GREATER THAN 2000 OHMS. IT WAS NOTED THERE WERE NO SIGNS OR SYMPTOMS RELATED TO THE EVENT AND THE PATIENT DID NOT REQUIRE HOSPITALIZATION. IT WAS STATED THAT EVEN THOUGH THE PATIENT HAD ABNORMAL IMPEDANCES THEY STILL HAD GOOD TREMOR CONTROL WHILE USING THE SAME ELECTRODE CONFIGURATION.

Description of Event or Problem · 1

IT WAS REPORTED THERE WERE IMPEDANCE MEASUREMENTS GREATER THAN 2,000 OHMS ON BIPOLAR PAIRS. IT WAS ALSO REPORTED THE ELECTRODE PAIR 1-3 HAD AN IMPEDANCE READING GREATER THAN 2,000 OHMS FOR THE LEFT IMPLANTABLE NEUROSTIMULATOR (INS) IN (B)(6) 2012. IT WAS NOTED THE RIGHT INS IMPEDANCE READINGS WERE GREATER THAN 2,000 OHMS FOR ALL BIPOLAR PAIRS. IT WAS ALSO STATED THE CURRENT MEASUREMENTS AT THE TIME OF REPORT AND THE PREVIOUS MEASUREMENTS WERE BETWEEN 9-12 MICROAMPS, WHICH IS POSSIBLY INDICATIVE OF AN OPEN CIRCUIT. IT WAS REPORTED IT HAD BECOME HARDER FOR THE PATIENT TO WRITE, WHICH HAD BEEN A PROBLEM "MORE SO LATELY." ADDITIONAL INFORMATION HAS BEEN REQUESTED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE FILED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194391 SOLETRA STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR MHY MEDTRONIC MED REL MEDTRONIC PUERTO RICO 7426

Patients

Seq Age Sex Outcome Treatment
1