FDA Adverse Event Malfunction Summary report: N

TROCHANTERIC NAIL KIT, STST GAMMA3 Ø11X180MM X 130°

MDR report key: 3092564 · Received May 3, 2013

Report

Report Number
0009610622-2013-00239
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 9, 2013
Report Date
April 10, 2013
Manufacturer
STRYKER OSTEOSYNTHESIS-KIEL
Product Code
HSB
PMA / PMN Number
K043431
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: MANUFACTURING OR MATERIAL FAULTS WERE NOT FOUND. APPEARANCE OF THREAD FLANKS AND TEXTURE OF SURFACES WITH CLEAR TRACES OF CROSS-THREADING INDICATE THE SET SCREW HAVING BEEN INSERTED UNDER SIGNIFICANT MISALIGNMENT BY IMMENSE FORCE. EVALUATION REVEALED EVIDENCE THAT THE EVENT IS NOT LINKED TO A DEFICIENCY OF THE DEVICES BUT IS RATHER RELATED TO USE ERROR.

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THE SURGEON, REPORTED TO THE SALES REP, THAT "WHEN PLACING THE GAMMA NAIL, WHEN THE SURGEON WANTED TO PUT THE SET SCREW IN PLACE, HE WAS UNABLE TO FULLY SET THIS SCREW. HE HAD TO REMOVE THE NAIL AND THE LAG SCREW, AND HE HAD TO PLACE ANOTHER NAIL. THERE WAS NO OTHER 130° NAIL AVAILABLE IN THIS HOSPITAL, SO HE HAD TO PUT A 125° NAIL. THE SURGEON SAID THIS WAS SUB OPTIMAL."

Description of Event or Problem · 1

THE SURGEON, REPORTED TO THE SALES REP, THAT "WHEN PLACING THE GAMMA NAIL, WHEN THE SURGEON WANTED TO PUT THE SET SCREW IN PLACE, HE WAS UNABLE TO FULLY SET THIS SCREW. HE HAD TO REMOVE THE NAIL AND THE LAG SCREW, AND HE HAD TO PLACE ANOTHER NAIL. THERE WAS NO OTHER 130° NAIL AVAILABLE IN THIS HOSPITAL, SO HE HAD TO PUT A 125° NAIL. THE SURGEON SAID THIS WAS SUB OPTIMAL."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194946 TROCHANTERIC NAIL KIT, STST GAMMA3 Ø11X180MM X 130° IMPLANT HSB STRYKER OSTEOSYNTHESIS-KIEL K241609

Patients

Seq Age Sex Outcome Treatment
1 Other