V40 COCR LFIT HEAD 40MM/+4
Report
- Report Number
- 0002249697-2013-01532
- Event Type
- Injury
- Date Received
- May 3, 2013
- Date of Event
- April 10, 2013
- Report Date
- April 10, 2013
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- LPH
- PMA / PMN Number
- K061434
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- PHYSICIAN
Narratives
WHEN COMPLETED, THE EVALUATION SUMMARY WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT. (B)(4): NOT RETURNED TO THE MANUFACTURER.
AN EVENT REGARDING ALTR INVOLVING A V40 COCR LFIT HEAD 40MM/+4 WAS REPORTED. THE EVENT WAS NOT CONFIRMED. A MATERIAL ANALYSIS REPORT CONCLUDED, ¿THE MACHINED TAPERED SURFACE OF THE V40 LFIT HEAD EXHIBITED EVIDENCE OF SCRATCHES FROM THE EXPLANTATION PROCESS AS WELL AS MICRO-MOTION DAMAGE AND CORROSION DEBRIS INDICATIVE OF A MECHANICALLY ASSISTED CORROSION PROCESS. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE V40 LFIT HEAD.¿ A DEVICE HISTORY REVIEW CONFIRMED ALL DEVICES ACCEPTED INTO FINISHED GOODS CONFORMED TO SPECIFICATION. A COMPLAINT HISTORY REVIEW CONFIRMED NO OTHER SIMILAR EVENTS FOR THE REPORTED LOT. THE EXACT CAUSE OF THE EVENT COULD NOT BE DETERMINED BECAUSE FURTHER INFORMATION IS NEEDED TO COMPLETE THE INVESTIGATION FOR DETERMINING THE ROOT CAUSE. NO FURTHER INVESTIGATION FOR THIS EVENT IS POSSIBLE AT THIS TIME.
PATIENT HAD ADVERSE LOCAL TISSUE REACTION. DR. (B)(6), REVISED PATIENT'S LEFT HIP. A 36 C-TAPER BIOLOX HEAD AND V40 TO C-TAPER ADAPTER.
PATIENT HAD ADVERSE LOCAL TISSUE REACTION. DR. (B)(6) REVISED PATIENT'S LEFT HIP. THE 36 C-TAPER BIOLOX HEAD AND V40 TO C-TAPER ADAPTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 194045 | V40 COCR LFIT HEAD 40MM/+4 | IMPLANT | LPH | STRYKER ORTHOPAEDICS-MAHWAH | MHK128 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention |