FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3092441 · Received May 3, 2013

Report

Report Number
1416980-2013-11255
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 10, 2013
Report Date
April 10, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING ONSITE INVESTIGATION OF (B)(4). THE ROOT CAUSE OF THE SAFETY CLAMP BEING OUT OF CALIBRATION IS UNKNOWN. TO CORRECT THE CONDITION, THE SAFETY CLAMP SHIM WAS REPLACED AND THE CLAMP WAS CALIBRATED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A SAFETY CLAMP OUT OF CALIBRATION. NO ADDITIONAL INFORMATION IS AVAILABLE.