FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3092441
·
Received May 3, 2013
Report
- Report Number
- 1416980-2013-11255
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- April 10, 2013
- Report Date
- April 10, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THIS IS AN ANCILLARY SERVICE EVENT OPENED DURING ONSITE INVESTIGATION OF (B)(4). THE ROOT CAUSE OF THE SAFETY CLAMP BEING OUT OF CALIBRATION IS UNKNOWN. TO CORRECT THE CONDITION, THE SAFETY CLAMP SHIM WAS REPLACED AND THE CLAMP WAS CALIBRATED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
DURING PRODUCT EVALUATION BY A SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE A SAFETY CLAMP OUT OF CALIBRATION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 194361 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |