FDA Adverse Event Malfunction Summary report: N

AUTOPLEX SYSTEM WITH VERTAPLEX HV W/O NEEDLES

MDR report key: 3092437 · Received May 3, 2013

Report

Report Number
0001811755-2013-01008
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 11, 2013
Report Date
April 11, 2013
Manufacturer
STRYKER INSTRUMENTS-PUERTO RICO
Product Code
JDZ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED WHEN THE DEVICES ARE RECEIVED AND AFTER THE QUALITY INVESTIGATION HAS BEEN COMPLETED. NOT RECEIVED BY MANUFACTURER.

Additional Manufacturer Narrative · 1

BASED ON THE INVESTIGATION FINDINGS, THE CLAIMED CONDITION WAS CONFIRMED. UPON RETURNED UNIT EVALUATION, CEMENT LEAKAGE WAS DETECTED BETWEEN THE LUER SPINDLE AND LUER HOUSING FROM THE EXTENSION TUBE ASSEMBLY. THE O-RING WAS PARTIALLY SHIFTED FROM THE GROOVE AND BROKEN CAUSING THE CEMENT TO LEAK.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE AUTOPLEX SYSTEM WAS BEING USED IN A VERTEBROPLASTY PROCEDURE WHEN THE CEMENT WOULD NOT GO DOWN THE NEEDLE INTO THE VERTEBRAL BODY, AND CEMENT WAS SEEN LEAKING BETWEEN THE TWIST CAP OF THE SYRINGE EXTENSION. A BACK-UP DEVICE WAS USED WHEN THE EVENT OCCURED FOR A SECOND TIME. WHEN THE THIRD SET WAS OPENED, THE CEMENT WAS ABLE TO BE INJECTED. THERE WERE NO PATIENT OR USER INJURIES, AND NO ADVERSE CONSEQUENCES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE AUTOPLEX SYSTEM WAS BEING USED IN A VERTEBROPLASTY PROCEDURE WHEN THE CEMENT WOULD NOT GO DOWN THE NEEDLE INTO THE VERTEBRAL BODY, AND CEMENT WAS SEEN LEAKING BETWEEN THE TWIST CAP OF THE SYRINGE EXTENSION. A BACK-UP DEVICE WAS USED WHEN THE EVENT OCCURRED FOR A SECOND TIME. WHEN THE THIRD SET WAS OPENED, THE CEMENT WAS ABLE TO BE INJECTED. THERE WERE NO PATIENT OR USER INJURIES, AND NO ADVERSE CONSEQUENCES.