FDA Adverse Event Malfunction Summary report: N

CLEARSTAR PUMP

MDR report key: 3092429 · Received May 3, 2013

Report

Report Number
1527460-2013-00043
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 4, 2013
Report Date
April 4, 2013
Manufacturer
ABBOTT NUTRITION - COLUMBUS
Product Code
LZH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE TESTING AND INVESTIGATION RESULTS DID NOT CONFIRM THE COMPLAINT. THE PUMP DELIVERED AT -0.7% ACCURACY, WHICH IS WITHIN SPECIFICATION.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE REPORTED, LIST NUMBER M771, IS AN ABBOTT PRODUCT THAT IS MARKETED INTERNATIONALLY, WHICH IS THE SAME OR SIMILAR TO A DEVICE, LIST NUMBER 55239, THAT IS MARKETED DOMESTICALLY. ABBOTT NUTRITION STANDARD PROCEDURE INCLUDES ATTEMPTING TO OBTAIN ALL REQUIRED INFORMATION FROM THE SOURCE.

Description of Event or Problem · 1

THE COMPLAINANT REPORTED AN UNDER-DELIVERY. THE INTENDED DELIVERY VOLUME WAS 600 ML AT A RATE OF 300 ML/HOUR FOR TWO HOURS; HOWEVER, THE ACTUAL DELIVERY VOLUME WAS 285 ML.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194307 CLEARSTAR PUMP LZH ABBOTT NUTRITION - COLUMBUS M771

Patients

Seq Age Sex Outcome Treatment
1