- Report Number
- 1416980-2013-11253
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Report Date
- April 12, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4). A BATCH REVIEW WAS CONDUCTED AND REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.EVALUATION SUMMARY:THE ACTUAL SAMPLE AND ONE COMPANION SAMPLE, FROM THE SAME LOT, WERE EVALUATED. VISUAL EXAMINATION OF THE ACTUAL SAMPLE NOTED THAT A SECTION OF THE MALE LUER SHAFT WAS MISSING AND WHITE STRESS MARKS WERE FOUND AROUND THE OPENING. THIS WAS LOCATED ON THE OPPOSITE SIDE OF THE MISSING SECTION. VISUAL EXAMINATION OF THE COMPANION SAMPLE NOTED NO SIGN OF PHYSICAL ABNORMALITY. HOWEVER, THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED.
(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED.
IT WAS REPORTED THAT AN INTERLINK SYSTEM SECONDARY MEDICATION SET LEAKED. DURING INFUSION, THE NURSE OBSERVED AN UNKNOWN MEDICATION LEAKING FROM THE LUER ATTACHED TO THE PRIMARY LINE. THE LUER WAS NOTED TO HAVE A ?1.5 CM LONG BY 1/3 DIAMETER? CHUNK MISSING; WITH A SOLID WHITE ACCENT. THE AMOUNT OF MEDICATION SPILLED IS UNKNOWN. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 195350 | ACCESS | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BAXTER HEALTHCARE - SINGAPORE | SR12125012 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN PRIMARY SET| UNKNOWN MEDICATION |