FDA Adverse Event Injury Summary report: N

STARCLOSE SE VASCULAR CLOSURE SYSTEM

MDR report key: 3092272 · Received May 3, 2013

Report

Report Number
2024168-2013-02748
Event Type
Injury
Date Received
May 3, 2013
Date of Event
April 11, 2013
Report Date
April 12, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P050007
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED EXPERIENCE WAS CONFIRMED. BASED ON VISUAL INSPECTION AND FUNCTIONAL TESTING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT-HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT AFTER A DIAGNOSTIC CORONARY ANGIOGRAM, ARTERIOTOMY CLOSURE OF THE COMMON FEMORAL ARTERY WAS ATTEMPTED USING A STARCLOSE SE DEVICE. REPORTEDLY, AS THE FLEX-GUIDE WAS INSERTED INTO THE EXCHANGE SHEATH, THE EXCHANGE SHEATH CUTTER WAS NOTICED TO BE BENT UPWARD. A GUIDE WIRE WAS INSERTED INTO THE EXCHANGE SHEATH. THE EXCHANGE SHEATH WAS REMOVED OVER THE GUIDE WIRE AND A SECOND STARCLOSE SE VASCULAR CLOSURE SYSTEM WAS USED TO ACHIEVE HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE AND NO REPORTED SIGNIFICANT CLINICAL DELAY IN THE PROCEDURE. THE OPERATOR WAS REPORTED TO BE TRAINED IN THE USE OF THE STARCLOSE SE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
195369 STARCLOSE SE VASCULAR CLOSURE SYSTEM IMPLANTABLE CLIP MGB AV-TEMECULA-CT 30315K1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6-FRENCH