FLOGARD
Report
- Report Number
- 1416980-2013-11237
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- April 11, 2013
- Report Date
- April 11, 2013
- Manufacturer
- SHARP CORPORATION OSAKA JAPAN
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH AN F-38 FAILURE WHEN THE PUMP WAS TURNED ON WAS CONFIRMED AND REPAIRED BY BAXTER SERVICE PERSONNEL ONSITE AT THE CUSTOMER FACILITY. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE DAMAGED FORCE-SENSING RESISTORS. THE FORCE-SENSING RESISTORS WERE REPLACED TO CORRECT THIS CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A FLO-GARD INFUSION PUMP WITH AN F-38 FAILURE WHEN THE PUMP WAS TURNED ON. IT IS UNKNOWN WHEN, OR IN WHICH CARE AREA, THIS EVENT OCCURRED. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE.