FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3092171 · Received May 3, 2013

Report

Report Number
1416980-2013-11229
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 14, 2013
Report Date
April 15, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER?S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN TUBING) ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, DURING DWELL 2 OF 5. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HP CYCLE POWER AND INFORMED THE HOME PATIENT (HP) THAT THE THERAPY HAD ENDED. THE HP DID NOT SAY IF THEY WOULD DO A MANUAL EXCHANGE OR SETUP THE HC WITH ALL NEW SUPPLIES. THE PATIENT WAS CONNECTED EITHER AT THE TIME OF THE ALARM OR OBSERVED AIR. DURING TROUBLESHOOTING, THERE WAS NOTHING FOUND THAT CAUSED OR CONTRIBUTED TO THE ALARM. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193740 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 45 YR HOME CHOICE