FDA Adverse Event Malfunction Summary report: N

NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE

MDR report key: 3092134 · Received May 3, 2013

Report

Report Number
0001811755-2013-01007
Event Type
Malfunction
Date Received
May 3, 2013
Date of Event
April 10, 2013
Report Date
April 10, 2013
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
FYD
PMA / PMN Number
K012991
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE TECHNICIAN REPLACED THE PLUG ON THE END OF THE POWER CORD, TESTED THE UNIT AND RETURNED THE UNIT TO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE POWER CORD ON THE NEPTUNE ROVER WAS SPARKING WHEN THE UNIT WAS BEING DOCKED. THERE WAS NO PATIENT INVOLVEMENT AND NO USER OR PATIENT INJURY AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194200 NEPTUNE ROVER WITH SMOKE EVAC AND POWER POLE APPARATUS, EXHAUST, SURGICAL FYD STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1