FDA Adverse Event
Malfunction
Summary report: N
HOMECHOICE
MDR report key: 3092117
·
Received May 3, 2013
Report
- Report Number
- 1416980-2013-11224
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- February 9, 2013
- Report Date
- April 15, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ADDITIONA RELEVANTL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
Description of Event or Problem · 1
A HIGH DRAIN ERROR 108 (NIGHT DRAIN #8) ALARM WAS IDENTIFIED IN THE LOG OF A RETURNED HOMECHOICE DEVICE. (A HIGH DRAIN ALARM IS INDICATIVE OF AN IIPV SITUATION. THIS ALARM INDICATES THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME IN STANDARD MODE.) THE ALARM OCCURRED ON (B)(6) 2013 20:30:16. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 194417 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 |