RESERVOIR 3ML
Report
- Report Number
- 3004209178-2013-93090
- Event Type
- Injury
- Date Received
- May 3, 2013
- Date of Event
- April 29, 2013
- Report Date
- April 29, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- FRN
- PMA / PMN Number
- K032005
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER
Narratives
INSPECTED THREE RANDOMLY SEALED RESERVOIRS. PERFORMED MANUAL PRIME AND HIGH PRESSURE TEST PER SPECIFICATIONS. RAN PRIMING IN INSULIN PUMP. RESERVOIRS PASSED PER SPECIFICATION. NO LEAKS OR O-RING DEFECTS WERE FOUND DURING TEST.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED THAT THE CUSTOMER HAD UNEXPLAINED LOW BLOOD GLUCOSE AND WAS HOSPITALIZED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS UNKNOWN. CALLER STATED THAT THE CUSTOMER'S RESERVOIR LEAKED INSULIN DURING NORMAL USED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193574 | RESERVOIR 3ML | RESERVOIR | FRN | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-332A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |