FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3092082
·
Received May 3, 2013
Report
- Report Number
- 3004209178-2013-93125
- Event Type
- Injury
- Date Received
- May 3, 2013
- Date of Event
- April 8, 2013
- Report Date
- April 9, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED ALL FUNCTIONAL TEST, INCLUDING PRIME, DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, AND EXCESSIVE NO DELIVERY ALARM TEST. NO MOTOR ERROR ALARM NOTED. UNIT HAD A SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT AND MISSING END CAP STICKER.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT HE HAD UNEXPLAINED LOW BLOOD GLUCOSE. CUSTOMER WENT TO THE EMERGENCY ROOM AND WAS HOSPITALIZED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE EMERGENCY ROOM VISIT WAS 53 MG/DL. CUSTOMER STATED THAT HE HAD MULTIPLE NO DELIVERY ALARMS AND MOTOR ERROR ALARMS. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193573 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR | Hospitalization |