FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3092064
·
Received May 3, 2013
Report
- Report Number
- 3004209178-2013-93107
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Date of Event
- April 7, 2013
- Report Date
- April 9, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARM MOTOR ERROR DURING THE BASIC OCCLUSION TEST, DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNIT WAS RECEIVED WITH MISSING END CAP STICKER. MOTOR PASSED MOTOR TEST.
Description of Event or Problem · 1
CUSTOMER REPORTED THAT SHE HAD HIGH BLOOD GLUCOSE; STATED THAT HER INSULIN PUMP DRIVE SUPPORT CAP IS STICKING OUT AND UNIT ALARMED WHILE TRYING TO PRIME. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193959 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |