FDA Adverse Event Death Summary report: N

ENDURANT II

MDR report key: 3092043 · Received May 3, 2013

Report

Report Number
2953200-2013-00815
Event Type
Death
Date Received
May 3, 2013
Date of Event
April 5, 2013
Report Date
April 5, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100020
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS: ENDOLEAK, VESSEL PERFORATION, DEATH. NO AORTIC NECK LENGTH, CONICAL PROXIMAL NECK, SEVERELY DISEASED AND FRAIL VESSELS. NO AORTIC NECK LENGTH. CONCLUSION: NO AORTIC NECK LENGTH, CONICAL PROXIMAL NECK, SEVERELY DISEASED AND FRAIL VESSELS. NO AORTIC NECK LENGTH.

Description of Event or Problem · 1

AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A CONICAL 5.7 CM IN DIAMETER ABDOMINAL AORTIC ANEURYSM THAT NARROWED TO 2.7 CM AND EXPANDED OUT TO APPROXIMATELY 4 CM IN DIAMETER. THE PATIENT WAS NOT A CANDIDATE FOR OPEN SURGICAL REPAIR. VESSEL MORPHOLOGY WAS REPORTED AS THERE WAS NO AORTIC NECK, THE VESSEL WAS SEVERELY DISEASED AND FRAIL, THE AORTIC NECK WAS 25 MM IN DIAMETER AT THE RENAL ARTERIES AND 10 MM BELOW THE RENAL ARTERIES THE AORTIC NECK WAS 29 MM IN DIAMETER. THERE WAS MILD THROMBOSIS THROUGHOUT THE VESSELS. THE STENT GRAFTS WERE IMPLANTED HOWEVER SINCE THERE WAS NO AORTIC NECK THERE WAS A LARGE PROXIMAL TYPE I ENDOLEAK. IT WAS REPORTED THAT THE PHYSICIAN ELECTED TO MODEL THE STENT GRAFT WITH A RELIANT BALLOON WITH NORMAL SALINE CONTRAST SOLUTION DURING THE MODELING OF THE STENT GRAFT THE VESSEL WAS PERFORATED WHERE THE ANEURYSM/AORTA NARROWED DOWN TO 2.7 CM IN DIAMETER, AND THE PATIENT'S BLOOD PRESSURE DROPPED. THE PHYSICIAN OPENED THE PATIENT AND CLAMPED THE AORTA ABOVE THE BIFURCATED STENT GRAFT TO GAIN BLOOD PRESSURE CONTROL HOWEVER, WHEN THE CLAMP WAS PLACED THE AORTA DISINTEGRATED. THE PHYSICIAN ATTEMPTED TO CLAMP HIGHER PROXIMALLY TO THE CELIAC ARTERY, BUT THE AORTA CONTINUED TO DISINTEGRATE. THE PHYSICIAN WAS UNABLE TO CONTROL THE BLEEDING AND THE PATIENT EXPIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194786 ENDURANT II SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND V03997534

Patients

Seq Age Sex Outcome Treatment
1 00083 YR Death