LINOX SMART SD 60/16

MDR report key: 3092004 · Received May 3, 2013

Report

Report Number
1028232-2013-01306
Event Type
Injury
Date Received
May 3, 2013
Date of Event
April 24, 2012
Report Date
April 22, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVY
PMA / PMN Number
P980023
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO A FUNCTIONAL ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. DURING THIS ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

THIS LEAD WAS EXPLANTED AND REPLACED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193921 LINOX SMART SD 60/16 ICD LEAD NVY BIOTRONIK SE & CO. KG 359065

Patients

Seq Age Sex Outcome Treatment
1 75 YR Hospitalization