OT VERIO PRO METER
Report
- Report Number
- 3008382007-2013-10063
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Report Date
- April 8, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- Removal / Correction Number
- 2013-IE-017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 (06/13/2013)-DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE TEST STRIPS WERE ALSO TESTED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE LAY USER/PATIENT'S PRODUCT(S) HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED IN THIS CASE HAVE PASSED ALL TESTING AND THE COMPLAINT WAS NOT CONFIRMED. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
ON (B)(6) 2013, THE LAY USER/PATIENT IN (B)(4) CONTACTED LIFESCAN (LFS) TO REPORT THE ONE TOUCH VERIO PRO METER WAS GIVING INACCURATE READINGS. THE PATIENT OBTAINED THE BLOOD GLUCOSE READING OF 93 MG/DL ON THE REPORTED METER AND A LABORATORY RESULT OF 70 MG/DL. THE PATIENT'S TESTING TECHNIQUE WAS CORRECT AND THE TEST STRIPS WERE IN GOOD CONDITION AND WITHIN OPENED EXPIRATION DATING. THE PATIENT DID NOT REPORT EXPERIENCING ANY SYMPTOMS OR MEDICAL ATTENTION. THE PATIENT DID NOT SUFFER ANY INJURY DUE TO THE REPORTED METER. HOWEVER, AS THE TEST RESULTS FELL OUTSIDE EXPECTED VALUES FOR METER TO LAB ACCURACY TESTING THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 193759 | OT VERIO PRO METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3363347 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR |