FDA Adverse Event
Malfunction
Summary report: N
DPM CENTRAL STATION
MDR report key: 3091887
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00545
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- October 1, 2012
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 080192
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REPRESENTATIVE IDENTIFIED AN ISSUE WITH THE FACILITIES' POWER SUPPLY OUTLET. SYSTEM WAS RETURNED TO USE USING A TEMPORARY POWER SUPPLY OUTLET.
Description of Event or Problem · 1
CUSTOMER REPORTED THE DPM CENTRAL STATION REPEATEDLY SHUT DOWN, WHICH MAY HAVE RESULTED IN A LOSS OF TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186824 | DPM CENTRAL STATION | CENTRAL MONITORING STATION | MHX | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |