Report Number
2221819-2013-01056
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
April 12, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REPRESENTATIVE CLEARED SYSTEM'S ERROR LOGS AND ADVISED THE CUSTOMER TO READJUST CONFIGURATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE PANORAMA CENTRAL STATION FROZE, WHICH MAY HAVE RESULTED IN LOSS OF TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188755 PANORAMA CENTRAL STATION CENTRAL MONITORIN STATION MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1