FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 3091820 · Received April 30, 2013

Report

Report Number
2221819-2013-00994
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
April 1, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REP REPLACED THE POWER SUPPLY AND THE PROBLEM RESOLVED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE PANORAMA CENTRAL STATION DISPLAY WOULD NOT TURN ON, WHICH MAY HAVE RESULTED IN LOSS OF TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188054 PANORAMA CENTRAL STATION CENTRAL MONITORING STATION MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1