FDA Adverse Event Malfunction Summary report: N

ACCUTORR V MONITOR

MDR report key: 3091814 · Received April 30, 2013

Report

Report Number
2221819-2013-01018
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
March 28, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
091068
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REPRESENTATIVE EVALUATED THE UNIT AND REPLACED THE CPU BOARD. UNIT CALIBRATED AND SAFETY TESTED TO FACTORY SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED AN ISSUE WITH THE ACCUTORR V MONITOR'S DISPLAY, WHICH MAY HAVE RESULTED IN LOSS OF PRIMARY MONITORING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
188053 ACCUTORR V MONITOR PATIENT MONITOR MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1