FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 3091633 · Received April 30, 2013

Report

Report Number
2221819-2013-00679
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
November 14, 2012
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

MINDRAY SERVICE REPS CLEARED ERROR LOGS AND REBOOTED SERVER AND COMMUNICATION WAS RESTORED.

Description of Event or Problem · 1

CUSTOMER REPORTED THE PANORAMA TELEPACK LOST COMMUNICATIONS WITH THE PANORAMA CENTRAL STATION WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
187175 PANORAMA CENTRAL STATION CENTRAL STATION MONITOR MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1