FDA Adverse Event Malfunction Summary report: N

PASSPORT 2 MONITOR

MDR report key: 3091629 · Received April 30, 2013

Report

Report Number
2221819-2013-00687
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
November 20, 2012
Report Date
April 30, 2012
Manufacturer
MINDRAY DS USA, INC.
Product Code
DSI
PMA / PMN Number
993531
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

MINDRAY SERVICE REPS FOUND BURNT COMPONENTS IN THE POWER SUPPLY. POWER WAS REPLACED, UNIT WAS CALIBRATED, SAFETY TESTED, AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED SMOKE WAS EMITTED FROM THE PASSPORT 2 MONITOR WHEN THE UNIT WAS POWERED ON. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186627 PASSPORT 2 MONITOR PATIENT MONITOR DSI MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1