FDA Adverse Event Malfunction Summary report: N

V SERIES MONITOR

MDR report key: 3091568 · Received April 30, 2013

Report

Report Number
2221819-2013-00409
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
December 6, 2012
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
102004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

MINDRAY SERVICE REPRESENTATIVES FOUND THE ECG ISSUES WERE SETTINGS RELATED AND RETURNED THE UNIT TO MINDRAY'S REPAIR CENTER FOR THE TOUCH SCREEN LOCK UP ISSUE. THE TOUCH SCREEN WAS REPLACED, SOFTWARE WAS UPGRADED, CALIBRATED AND SAFETY TESTED TO FACTORY SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE V SERIES MONITOR TOUCH SCREEN LOCKED UP ISSUES WITH THE ECG, WHICH MAY HAVE RESULTED IN A POSSIBLE LOSS OF PRIMARY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
187384 V SERIES MONITOR PATIENT MONITOR MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1