FDA Adverse Event
Malfunction
Summary report: N
V SERIES MONITOR
MDR report key: 3091568
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00409
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- December 6, 2012
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 102004
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REPRESENTATIVES FOUND THE ECG ISSUES WERE SETTINGS RELATED AND RETURNED THE UNIT TO MINDRAY'S REPAIR CENTER FOR THE TOUCH SCREEN LOCK UP ISSUE. THE TOUCH SCREEN WAS REPLACED, SOFTWARE WAS UPGRADED, CALIBRATED AND SAFETY TESTED TO FACTORY SPECIFICATIONS.
Description of Event or Problem · 1
CUSTOMER REPORTED THE V SERIES MONITOR TOUCH SCREEN LOCKED UP ISSUES WITH THE ECG, WHICH MAY HAVE RESULTED IN A POSSIBLE LOSS OF PRIMARY MONITORING. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 187384 | V SERIES MONITOR | PATIENT MONITOR | MHX | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |