Summary report: N
PASSPORT 2 MONITOR
MDR report key: 3091549
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00728
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- October 26, 2012
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- DSI
- PMA / PMN Number
- 020550
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REPRESENTATIVE REPLACED THE SPO2 BOARD AND FAN AND FOUND THE UNIT WAS PACKED WITH LINT WHICH CAUSED OVERHEATING.
Description of Event or Problem · 1
CUSTOMER REPORTED AN ISSUE WITH THE PASSPORT 2 MONITOR, WHICH MAY HAVE RESULTED IN AN INTERMITTENT LOSS OF SPO2 MONITORING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186482 | PASSPORT 2 MONITOR | PATIENT MONITOR | DSI | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|