FDA Adverse Event
Malfunction
Summary report: N
PASSPORT 2 MONITOR
MDR report key: 3091520
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00343
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- August 22, 2012
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 993531
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REPRESENTATIVE EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACEMENT OF THE SPO2 CABLE, REPROGRAMMING THE CPU BOARD, AND RESEATING THE CABLES. THE UNIT WAS TESTED TO FACTORY'S SPECIFICATIONS. IF FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED INTERMITTENT SPO2 READING ON THE PASSPORT 2 MONITOR, WHICH MAY HAVE RESULTED IN LOSS OF SPO2 MONITORING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186786 | PASSPORT 2 MONITOR | PATIENT MONITOR | MHX | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |