FDA Adverse Event
Malfunction
Summary report: N
GAS MODULE II
MDR report key: 3091513
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00462
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- December 19, 2012
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- BZK
- PMA / PMN Number
- 062754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REPRESENTATIVES REPLACED THE UNIT BENCH ASSEMBLY. UNIT WAS CALIBRATED AND TESTED TO PRODUCT SPECIFICATION.
Description of Event or Problem · 1
CUSTOMER REPORTED NO O2 READINGS ON THE GAS MODULE II WHICH MAY HAVE RESULTED IN A LOSS OF O2 MONITORING. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186441 | GAS MODULE II | GAS MONITOR | BZK | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |