FDA Adverse Event Injury Summary report: N

LCP DIST-TIBPL3.5 ANTERO-LAT R 13HO L184

MDR report key: 3091470 · Received May 3, 2013

Report

Report Number
8030965-2013-02108
Event Type
Injury
Date Received
May 3, 2013
Date of Event
March 4, 2013
Report Date
April 8, 2013
Manufacturer
SYNTHES GMBH
Product Code
KTT
PMA / PMN Number
K000684
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS FAR AS VISIBLE ON THE X-RAY IMAGES THE PLATE WAS FIXED WITH THIRTEEN SCREWS AND ONE ADDITIONAL SCREW WAS SET NEAR THE PLATE. THE BREAKAGE OF THE INVESTIGATED LCP WAS DUE TO DYNAMIC BENDING WHICH LED TO THE MATERIAL FATIGUE. BASED ON THE TOPOGRAPHY OF THE FRACTURE SURFACES WE CAN CONCLUDE THAT THE INVESTIGATED IMPLANT HAD TO ABSORB AND NEUTRALIZE FORCES THAT MAY HAVE BEEN GREATER THAN THE TOLERABLE LOAD. POSTOPERATIVE ACTIVITIES OF THE PATIENT MAY HAVE PLAYED A CERTAIN ROLE. A FAILURE RESULTING FROM EITHER A MATERIAL DEFECT OR THE MANUFACTURING PROCESS CAN BE EXCLUDED. THIS COMPLAINT IS INVALID. ONE OF THE DATES OF THE X-RAYS WAS REPORTED INCORRECTLY ON THE INITIAL REPORT. IT HAS BEEN CORRECTED HEREIN. PLACEHOLDER.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. THE INVESTIGATION COULD NOT COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED.

Description of Event or Problem · 1

PLATE BROKE AT A SCREW HOLE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: ON (B)(6) 2012, A PATIENT WAS IMPLANTED WITH THE LCP ALDTP AND LAG SCREW FOR A RIGHT TIBIA SHAFT FRACTURE. A PTB BRACE WAS APPLIED ON (B)(6) 2012. ON (B)(6) 2013, THE SURGEON FOUND METAL BROKEN THROUGH X-RAY. THE SURGEON REMOVED THE BROKEN PLATE AND INSERTED A NAIL ON (B)(6) 2013. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
193955 LCP DIST-TIBPL3.5 ANTERO-LAT R 13HO L184 KTT SYNTHES GMBH 7622577

Patients

Seq Age Sex Outcome Treatment
1 81 YR Required Intervention