FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 3091467 · Received April 30, 2013

Report

Report Number
2221819-2013-00984
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
August 23, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENT OF THE SYSTEM'S DISCLOSURE DRIVE. SYSTEM WAS TESTED TO FACTORY'S SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE PANORAMA CENTRAL STATION HAD AN ERROR MESSAGE, WHICH MAY HAVE RESULTED IN LOSS OF TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186695 PANORAMA CENTRAL STATION CENTRAL STATION MONITOR MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1