FDA Adverse Event Malfunction Summary report: N

GAS MODULE II

MDR report key: 3091460 · Received April 30, 2013

Report

Report Number
2221819-2013-00358
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
August 27, 2012
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
BZK
PMA / PMN Number
062754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACING THE GAS BENCH AND CALIBRATING THE UNIT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE GAS MODULE II WON'T TAKE GAS READINGS. NO PT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
187029 GAS MODULE II GAS MODULE BZK MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1