FDA Adverse Event Malfunction Summary report: N

PASSPORT V MONITOR

MDR report key: 3091449 · Received April 30, 2013

Report

Report Number
2221819-2013-00424
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
September 10, 2012
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC
Product Code
MHX
PMA / PMN Number
091834
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACEMENT OF THE ECG PARAMETER MODULE. THE UNIT WAS TESTED TO FACTORY'S SPECS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED NO ECG READINGS ON THE PASSPORT V MONITOR, WHICH MAY HAVE RESULTED IN LOSS OF ECG MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186684 PASSPORT V MONITOR PATIENT MONITOR MHX MINDRAY DS USA, INC

Patients

Seq Age Sex Outcome Treatment
1