Narratives
Additional Manufacturer Narrative · 1
THE COMPANY REPRESENTATIVE EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACEMENT OF THE ECG PARAMETER MODULE. THE UNIT WAS TESTED TO FACTORY'S SPECS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED NO ECG READINGS ON THE PASSPORT V MONITOR, WHICH MAY HAVE RESULTED IN LOSS OF ECG MONITORING. NO PT INJURY WAS REPORTED.