FDA Adverse Event Injury Summary report: N

2.4MM TI VA LOCKING SCREW STARDRIVE 18MM

MDR report key: 3091440 · Received May 3, 2013

Report

Report Number
2520274-2013-02335
Event Type
Injury
Date Received
May 3, 2013
Report Date
April 8, 2013
Manufacturer
SYNTHES USA
Product Code
HWC
PMA / PMN Number
K103243
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THIS SCREW IS WHOLE AND INTACT. THE DIAMETER OF 2.38MM, STARDRIVE, THREADED HEAD, FLUTES, AND 18.38MM LENGTH CONFIRM THIS AS A 04.210.118. THE DRIVE IS IN GOOD CONDITION. THE TOP OF THE HEAD IS ALSO IN GOOD CONDITION. THE HEAD THREADS HAVE BEEN LIGHTLY DAMAGED IN THE FORM OF COMPRESSION OF THE MAJOR DIAMETER INTO A T. THIS DAMAGE AFFECTS PROFILE. THE FIRST THREE SHAFT THREADS HAVE THE ANODIZE COATING REMOVED. THE REMAINDER OF THE SHAFT THREADS ARE IN GOOD CONDITION, AS ARE THE FLUTES AND TIP. THE DIMENSIONS THAT COULD BE CHECKED WERE FOUND ACCEPTABLE. BECAUSE OF DAMAGE INCURRED, AND ALSO BECAUSE NO LOT NUMBER WAS PROVIDED, A FINITE EVALUATION COULD NOT BE PERFORMED.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. PLACEHOLDER.

Description of Event or Problem · 1

A VOLAR PLATE WAS IMPLANTED ON AN UNKNOWN DATE. ON AN UNKNOWN DATE, THE PLATE PULLED AWAY FROM THE BONE CAUSING THE PATIENT PAIN. THE SURGEON PERFORMED A HARDWARE REMOVAL OF THE PLATE AND SEVEN SCREWS ON (B)(6) 2013. A LARGER PLATE AND NINE SCREWS WERE IMPLANTED INSTEAD. THE SURGEON REPORTED STRONGER FIXATION WITH THE NEW HARDWARE. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 3 OF 8 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
194104 2.4MM TI VA LOCKING SCREW STARDRIVE 18MM HWC SYNTHES USA

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention