ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-05825
- Event Type
- Malfunction
- Date Received
- May 3, 2013
- Report Date
- April 9, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE PUMP¿S HISTORY SHOWED EVIDENCE OF REBOOTS ON THE REPORTED FAILURE DATE. THE PUMP WAS RETURNED WITH A CRACKED BATTERY COMPARTMENT AND AN UNDAMAGED BATTERY CAP; HOWEVER THE CAP WAS UNABLE TO TIGHTEN ON SECURELY. A TEST CAP WAS USED AND WAS ABLE TO TIGHTEN ON FULLY. THE PUMP POWERED ON AND DISPLAYED THE VERIFY SCREEN. THE PUMP WAS TESTED FOR 24 HOURS WITH NO POWER INTERRUPTION. THE PUMP COVER WAS REMOVED AND NO INTERNAL MOISTURE OR DAMAGE WAS FOUND TO THE POWER CIRCUIT OR BATTERY FLEX. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DIM AND DISCOLORED DISPLAY SCREEN. A TEST SCREEN WAS INSTALLED AND DISPLAYED NORMALLY. ADDITIONALLY, THE UP ARROW AND CONTRAST KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE DURING TESTING. THE KEYPAD COVER WAS REMOVED; CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS AND THE UP ARROW AND CONTRAST KEYPAD BUTTONS WERE MISALIGNED.
THE PUMP HAS BEEN RETURNED TO ANIMAS BUT EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
ON (B)(6) 2013, THE PATIENT CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 194593 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |