FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION

MDR report key: 3091294 · Received April 30, 2013

Report

Report Number
2221819-2013-00845
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
January 29, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
031760
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

MINDRAY SERVICE REP EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACING THE COOLING FAN AND RADIO TRANSCEIVER BOARD. THE UNIT WAS TESTED TO FACTORY'S SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.

Description of Event or Problem · 1

CUSTOMER REPORTED AN ISSUE WITH THE PANORAMA CENTRAL STATION'S TRANSMITTER WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186448 PANORAMA CENTRAL STATION PATIENT MONITOR MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1