FDA Adverse Event
Malfunction
Summary report: N
PANORAMA CENTRAL STATION
MDR report key: 3091294
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00845
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- January 29, 2013
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- MHX
- PMA / PMN Number
- 031760
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REP EVALUATED THE UNIT. CORRECTIONS INCLUDED REPLACING THE COOLING FAN AND RADIO TRANSCEIVER BOARD. THE UNIT WAS TESTED TO FACTORY'S SPECIFICATIONS. IT FUNCTIONED NORMALLY AND WAS RETURNED TO USE.
Description of Event or Problem · 1
CUSTOMER REPORTED AN ISSUE WITH THE PANORAMA CENTRAL STATION'S TRANSMITTER WHICH MAY HAVE RESULTED IN A LOSS OF AMBULATORY TELEMETRY MONITORING. NO PATIENT INJURY WAS REPORTED.