FDA Adverse Event Malfunction Summary report: N

GAS MODULE 3 MONITOR

MDR report key: 3091283 · Received April 30, 2013

Report

Report Number
2221819-2013-00524
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
February 4, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
BZK
PMA / PMN Number
062754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

MINDRAY SERVICE REPS REPLACED THE MOISTURE TRAP AND RECALIBRATED TO FACTORY SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE GAS MODULE DISPLAYED INACCURATE O2 AND N2O READINGS. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
186734 GAS MODULE 3 MONITOR GAS MODULE BZK MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1