FDA Adverse Event
Malfunction
Summary report: N
GAS MODULE 3 MONITOR
MDR report key: 3091283
·
Received April 30, 2013
Report
- Report Number
- 2221819-2013-00524
- Event Type
- Malfunction
- Date Received
- April 30, 2013
- Date of Event
- February 4, 2013
- Report Date
- April 30, 2013
- Manufacturer
- MINDRAY DS USA, INC.
- Product Code
- BZK
- PMA / PMN Number
- 062754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
MINDRAY SERVICE REPS REPLACED THE MOISTURE TRAP AND RECALIBRATED TO FACTORY SPECIFICATIONS.
Description of Event or Problem · 1
CUSTOMER REPORTED THE GAS MODULE DISPLAYED INACCURATE O2 AND N2O READINGS. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 186734 | GAS MODULE 3 MONITOR | GAS MODULE | BZK | MINDRAY DS USA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |