FDA Adverse Event Malfunction Summary report: N

PASSPORT 2 MONITOR

MDR report key: 3091195 · Received April 30, 2013

Report

Report Number
2221819-2013-00754
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
February 26, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
MHX
PMA / PMN Number
020550
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REP EVALUATED THE SYS. REPLACED MEMORY CHIP, AND UPGRADED TO SYSTEM SOFTWARE. RESEATED CABLES. CALIBRATED AND SAFELY TESTED UNIT TO FACTORY SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE PASSPORT 2 MONITOR'S PORT WAS NOT RECOGNIZING THE GAS MODULE WHICH MAY HAVE RESULTED IN LOSS OF GAS MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
187322 PASSPORT 2 MONITOR PATIENT MONITOR MHX MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1