FDA Adverse Event Malfunction Summary report: N

GAS MODULE II MONITOR

MDR report key: 3091129 · Received April 30, 2013

Report

Report Number
2221819-2013-00923
Event Type
Malfunction
Date Received
April 30, 2013
Date of Event
March 13, 2013
Report Date
April 30, 2013
Manufacturer
MINDRAY DS USA, INC.
Product Code
BZK
PMA / PMN Number
062754
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

SERVICE REPRESENTATIVES REPLACED BROKEN O2 CONNECTOR. REPLACED THE SMALL DIAMETER TUBING BETWEEN THE O2 FILTER AND THE O2 SENSOR. PERFORMED THE RECOMMENDED 12 AND 24 MONTH PREVENTIVE MAINTENANCE. TESTED, CALIBRATED AND SAFETY CHECKED UNIT TO FACTORY SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED LOW FLOW GAS READINGS FROM THE GAS MODULE. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
187755 GAS MODULE II MONITOR GAS MODULE BZK MINDRAY DS USA, INC.

Patients

Seq Age Sex Outcome Treatment
1